PV Bulletin

Ventilateur Newport HT70 et Newport HT70 Plus – Medtronic France SAS

Medtronic is recalling certain serial numbers of Newport™ HT70 and HT70 Plus ventilators due to potential failure scenarios.

Primary source: Consultez le courrier de la société Medtronic France SAS de juin 2025 (05/06/2025)

Translated From French

PV Bulletin does not take any responsibility for translation errors or omissions, though it has used the best available translation engine.

PV Impact Brief

Urgency: HighConfidence: high

What changed

Medtronic has initiated a recall of specific serial numbers of the Newport HT70 and HT70 Plus ventilators and associated spare parts, recommending the cessation of clinical use of these devices.

Why it matters

The failure of these ventilators can result in serious health risks, including respiratory failure and potential fatalities, necessitating immediate action from healthcare providers to safeguard patient well-being.

Action needed

Healthcare facilities using the affected Newport HT70 and HT70 Plus ventilators need to cease their use and evaluate the risk for their patients, following established protocols for device monitoring and replacement.

Relevant for
Regulatory Intelligence Lead
Processes impacted
Signal Management
Owner

Regulatory Intelligence Lead

Review cadence

Review in the next regulatory intelligence cycle.

Evidence and confidence

Confidence: highSource updated: Sep 3, 2026

Full briefing

Practical implication

Healthcare facilities using the affected Newport HT70 and HT70 Plus ventilators need to cease their use and evaluate the risk for their patients, following established protocols for device monitoring and replacement.

Medtronic France SAS is recalling specific serial numbers of the Newport HT70 and HT70 Plus ventilators due to identified issues that could lead to serious incidents or fatalities. This recall stems from investigations into customer complaints revealing problems with circuit board capacitors that could lead to ventilator failure or failure to sound alarms in critical situations.

What changed: Medtronic has initiated a recall of specific serial numbers of the Newport HT70 and HT70 Plus ventilators and associated spare parts, recommending the cessation of clinical use of these devices.

Why it matters: The failure of these ventilators can result in serious health risks, including respiratory failure and potential fatalities, necessitating immediate action from healthcare providers to safeguard patient well-being.

Practical implication: Healthcare facilities using the affected Newport HT70 and HT70 Plus ventilators need to cease their use and evaluate the risk for their patients, following established protocols for device monitoring and replacement.

View original source

  • Source: ANSM (France) - Safety information feed
  • Published at source: 2026-09-03T00:00:00.000Z
  • Discovered by monitor: 2026-09-03T21:47:55.789Z

Published from the regulator source ingestion pipeline after PV impact triage.

Related updates for your role

publicFDA · USA · Updated Sep 4, 2026

FDA Grants Accelerated Approval to Etcamah for Advanced Breast Cancer

Urgency: High

Regulatory teams must prepare for updating labeling to include details on the new indication and required diagnostic methods, along with planning for the confirmatory studies mandated by the FDA.

Signal ManagementLabelingRegulatory Intelligence Lead
SOP Impact2 SOPs
ConfidenceHigh
Relevant for: Regulatory Intelligence Lead
verifiedRegulatory action: Regulators
Read impact briefarrow_forward
publicMHRA · UK · Updated Sep 4, 2026

Change to the definition of regulated activity with children from 1 September 2026

Urgency: High

Organizations must re-evaluate roles related to children's activities and prepare to request Enhanced DBS checks including access to the Children’s Barred List check starting from 1 September 2026.

Country ComplianceRegulatory Intelligence Lead
SOP Impact1 SOPs
ConfidenceHigh
Relevant for: Regulatory Intelligence Lead
verifiedRegulatory action: Regulators
Read impact briefarrow_forward
publicUK Health Security Agency · UK · Updated Sep 7, 2026

Low Flu Vaccination Rates Among Primary School Children

Urgency: High

Parents are reminded to complete and return flu vaccination consent forms promptly as vaccination sessions in schools are set to start this week, aiming for completion by late November.

Signal ManagementPV Quality Lead
SOP Impact1 SOPs
ConfidenceHigh
Relevant for: PV Quality Lead, Medical Safety Lead
verifiedRegulatory action: Regulators
Read impact briefarrow_forward
publicANSM (France) · France · Updated Sep 7, 2026Translated From French

Recall of Gore Viabahn Endoprosthesis with Bioactive Heparin Coating

Urgency: High

Health establishments receiving affected devices must be notified, and protocols established to ensure compliance with the recall.

Signal ManagementQPPV
SOP Impact1 SOPs
ConfidenceHigh
Relevant for: QPPV, Medical Safety Lead
verifiedRegulatory action: Regulators
Read impact briefarrow_forward